Item Description

Ziostand

Make an Offer

The Respironics Alice 5 Diagnostic Sleep System includes a desktop computer installed with Windows 10 and Alice G3 software, LCD monitor, mouse, keyboard, amplifier/base station, head box/input box, and interconnecting cables. It is certified to proper operating specifications and in excellent condition.


Respironics Alice 5 Diagnostic Sleep System Features
Fully integrated sleep laboratory system over a network cable or wireless.
Expanded channel capability to 55 total channels.
High quality ECG with six channels and Pulse Transit Time (PTT).
Real time impedance display.
Patient Communication/Intercom is accomplished over your network cable.
Respironics Alice 5 Diagnostic Sleep System Specifications
Total Channels: 55.
Analog Inputs: 12 total, 10 in the base station, 2 in the headbox.
Neurological Channels Specifications:
Number of Channels: 26 (10 left, 9 right, 2 EOG, 5 EMG) 2 reference.
Input Impedance: 1.66 Mohm per Electrode 3.33 differential.
Bandwith: 0.32 Hz to 106 Hz.
Input Signal Range: +/- 3.3 mV.
Digital Resolution: 16 bits.
Sample Rate: 2000 Hz.
Max Storage Rate: 200 Hz.
Electrode Impedance Testing: 100 Hz square wave between patient electrodes.
Calibration Signal: Signal injection, 1 Hz, 98 µVpp.
Communication Interfaces: Signals sent from Patient Interface to Headbox through sensor cables. Data is sampled and sent to the Base Station where it is stored on a disk or sent through an ethernet connection to a Host PC.
During Analysis: 0.5 to 100 Hz low pass filter 50 or 60 Hz Notch filter Anti-aliasing filter.
Other Channel Types:
(2) Airflow: Thermal and Pressure Flow.
(1) Body Position.
(2) Actigraphy.
(6) ECG.
(1) Snore Mic.
System Physical Characteristics:
Main Unit:
14" x 5" x 12.5" (35.6 cm x 12.7 cm x 31.7 cm).
Weight: 7.8 lb (3.5 kg).
Headbox:
10" x 4" x 2" (25.4 cm x 10.2 cm x 5.1 cm).
Weight: 1.6 lb (0.72 kg).
System Power Requirements:
Power to Main Unit: 6.25 VDC, 3 to 6 Amps.
Power Supply: MW116 Power Supply 100 - 240 VAC, 50 - 60 Hz, 0.5 - 1.5 A.