Toshiba PLT 1005BT Linear Probe Ultrasound Repair
Problem: Image drop out
Solution: replace crystal
Knowledge point
One may be facing several types of problems while using the ultrasound machine probes. They may be one or more of the following:
1. Poor quality image: Poor image may be due to
*Cracked or air bubbles on lens
*Connector issue
*Crystal damage
*Tears on the sheath
2. Physical damage or broken housing: Physical damage may be due improper handling, dropping of probe, poor storage etc.
3. Noise: Noise may be due to broken cable, damaged / broken pins in connector or electrical interference.
*Breakage of probe connector pins may be caused by improper insertion or connection.
*Damage to cable occurs due to cables being ridden by some wheels or closure of drawers without proper placement of the probe inside, etc.
4. No image: If there is image being received, it could be a result of probe cable – connector breakage, probe compatibility issue or software issue.