Here is a Bio-Rad QX200 Droplet Digital PCR system with AutoDG Automated Droplet Generator, PX1 Plate Sealer and C1000 Touch Thermal Cycler Unit3
Predecessor to:
Current model
Condition:
This Bio-Rad QX200 Droplet Digital PCR system is in good working condition.
Please see the exact video and photos below.
Testing we performed:
The Bio-Rad QX200 Droplet Digital PCR system passed all tests and was found to be fully functional:
Power: PASSED
Communication with software: PASSED
Prepared test sample using AutoDG: PASSED
Sample cover open/close: PASSED
Sample detection: PASSED
Fluid/Waste level sensors: PASSED
Fluid/Waste door sensors:: PASSED
Software controls: PASSED
Flush System: PASSED
Prime: PASSED
Ran test sample in QX200: PASSED
Tested C1000 through temperature range: PASSED
Sealed plate using PX1 PCR plate Sealer: PASSED
System consists of:
Bio-Rad QX200 Droplet Reader
Bio-Rad AutoDG Automated Droplet Generator
PX1 Plate Sealer
C1000 Touch Thermal Cycler
Computer with Bio-Rad software
Communication and power cables.
Configuration and Setup:
QX200 Droplet Reader
Precision: +/- 10%
Linear dynamic range: 5 orders of magnitude
Capacity: 1–96 samples
Droplets per 96-well plate: approximately 1,500,000
Sample illumination: light-emitting diodes
Sample detection: multipixel photon counter
Detection channels: FAM (EvaGreen), HEX (VIC)
Dimensions (W x D x H): 66 x 52 x 29 cm (26 x 20 x 11 in)
AutoDG Automated Droplet Generator
Starting sample size: 20 ul
Capacity: 1–96 samples/run
Droplets per sample: 20,000
Dimensions (W x D x H): 66 x 56 x 66 cm (26 x 22 x 26 in)
Weight: 100 lb (45.4 kg)
C1000 Touch Thermal Cycler
Input power: Up to 850 W maximum
Frequency: 50-60 Hz, single phase
Memory: >1,000 typical programs onboard
Temperature control modes: calculated and block
Programming options: step-based graphical and automatic
Reporting: exportable run logs, system error logs
Instant incubation
Dimensions (W x D x H): 33 x 45.5 x 20 cm (13 x 18 x 8 in)
Weight: 10.5 kg (23 lb)
PX1 PCR Plate Sealer
Heat up time: <3 min
Temperature range: 100-190 DegC
Time range: 0.5–10 s
Dimensions (W x D x H): 19.2 x 24.8 x 21.9 cm (7.6 x 9.8 x 8.6 in)
Weight: 7.3 kg (16 lb)
Power requirements: 100-240 VAC, 50/60 Hz 840 W
Video Description:
This video below gives an overview of Bio-Rad QX200 Droplet Digital PCR system with AutoDG Droplet Generator, PX1 Plate Sealer and C1000 Touch Thermal Cycler, and also demonstrates its functionality.
TO VIEW ADDITIONAL PHOTOS, VIDEOS AND VERY DETAILED INFORMATION SUCH AS TEST STEPS PERFORMED, CONFIGURATION, ETC., PLEASE VISIT OUR WEB SITE. TO GET THERE PLEASE CLICK ON “WEBSTORE”; TO YOUR RIGHT AND YOU WILL SEE LINK TO OUR SITE IN THE MIDDLE OF THE SCREEN. THEN USE LAST 5 DIGITS OF MPN NUMBER IN THE SEARCH FIELD TO LOCATE YOUR ITEM.